EYC THM50X Temperature & Humidity Transmitter
Application:
Semiconductor / Electronics / Chemical industry / Pharmaceutical / Steel industry / Biotechnology industry / Industrial process monitoring / Air conditioning / Cleanrooms / Laboratories / Weather stations / Hospitals / Buildings / Dryers / Factories / Storage rooms / Environmental ventilation control / Environmental test chambers / Greenhouses
Features:
● Aluminum alloy casing for harsh environments.
● Direct key operation for temperature/humidity adjustment, RS-485 settings, and unit switching.
● High-accuracy temperature and humidity measurement with fast response and quick recovery after condensation.
● RS-485 communication for configuring measurement units, ranges, output type, and station address.
● DATA LOG function capable of recording up to 65,535 entries for analysis.
● Measures extremely low humidity, down to 0.1% RH.
● Multiple physical measurement units switchable: %RH, °C, mbar, g/kg, g/m³, kJ/kg, AH%.
*AH% = the actual amount of water vapor in the air, defined with the saturated value at 25 °C as the reference. That is, the maximum water content at 25 °C is taken as 100%, and AH% represents the percentage calculated based on the corresponding absolute humidity (AH).
| Temperature | Actual Absolute Humidity (Measured) | Absolute Humidity Percentage (AH%) |
| 100°C | 4.6 g/m³ | (4.6 / 23) x 100% = 20% |
| 100°C | 1.15 g/m³ | (1.15 / 23) x 100% = 5% |
*Note: The saturated humidity at 25 °C is 23.03 g/m³.
Specification:
Input |
|
Signal type |
Capacitive humidity sensor & Pt100 Class A |
Measuring range |
Wall type:-40°C ... +80°C |
Output |
|
Output |
4 ... 20 mA / 0 ... 10 V / RS-485 |
Signal connection |
3-wire |
Warm-up time |
< 30 sec |
Response time (at 25°C) |
t90≦ 20 sec (S.S. Metal grid filter with mesh) |
Load resistance |
Current output:≦500 Ω |
Output calibration (Adjustment range) |
Software / Button |
Accuracy |
|
Temperature |
±0.1°C±0.002°C×tactual |
Humidity (0 ... 90%) |
Nonlinear error:±1%RH |
Humidity (90 ... 100%) |
±2%RH |
Temp. influence |
0.05%RH/°C |
Communication |
|
Comm. Interface & Protocol |
RS-485 Modbus RTU |
RS-485 Baud Rate |
9600、19200、38400、57600、115200 bps |
Environmental |
|
Measuring medium |
Air |
Temp. Operating (Housing) |
-20 ... +60°C |
Storage temperature |
-20 ... +60°C |
Probe pressure resistance |
4 bar |
Display unit |
°C / %RH / dp°C / AH% |
Electrical |
|
Power supply |
DC 24 V±10% |
Current consumption |
DC 24 V:60 mA |
Electrical connection |
M12 8P connector |
Installation |
|
Installation |
Wall / Remote |
Display |
|
Display readout |
Temperature/Humidity (initial value setting) |
Decimal point |
Button |
Sampling time |
1 cycle/sec |
Unit selection |
Temperature(T) / Humidity(H) / Dew Point(D) / Frost Point(F) / Wet-bulb Temperature(W) / |
Response time |
0 ... 100 |
Certification |
|
Certification |
CE |
Protection |
|
IP rating |
IP65 (Body) |
Electrical protection |
■ Reverse polarity ■ Over-voltage |
Material |
|
Housing |
Aluminum alloy |
Probe |
Stainless steel |
Remote probe cable |
TEFLON |
Weight |
THM501:170 g THM503:370 g |